Analog Output System for the Myomonitor®

The Myomonitor IV Analog Output System enables the integration of 16 full-bandwidth data channels from the Myomonitor IV Wireless EMG System into any analog data acquisition system. It uses a precision waveform generator with a low-noise filter to recreate the high-fidelity signals detected by Delsys Sensors.

Analog Output Device

Specifically designed for:

  • Gait Laboratories
  • Biomechanics Laboratories
  • Exercise Physiology Research
  • Ergonomic Studies
  • Robotics & HCID Interfacing

Features:

  • 16 channels, 4 kHz sampling
  • BNC & Mass terminations
  • +/- 5 V Range
  • Low noise, full signal conditioning
  • 50 ms monitored latency
 
Delsys Data Acquisition
 
Integrated Data Acquisition

Typical Laboratory Setup

Laboratory Set-UP

Analog Output System Components

Analog Output System Components

Analog Filter Module Specifications*

Filter Topology Butterworth, 8 pole
Bandwidth DC-500 Hz, +/- 10%
Voltage Range +/- 5 V
Setting Time <250 us
Baseline Noise 0.250 mV (RMS,RTO)
Input: Output Error (SNR) <2%
Power Supply 9 VDC
Dimensions 205 mm x 108 mm x 57 mm
Mass 300 g
Output Connections BNC, SCSI-68 (National Instruments)
Output Impedance <50mΩ ,buffered

NI PCI-6723 Waveform Generator Specifications

Output Range +/- 5 V
Resolution 13 Bits
Number of Channels 16
Update Rate 4000 samples/sec
PC Bus Requirement PCI
Dimensions 174 mm x98 mm
Operating Temperature 6-50 °C
Humidity 5-90% RH, non-condensing
Output Connections VHDCI (National Instruments)
 
*All specifications are subject to change without notice. Specifications include the software configuration of the device.