Analog Output System for the Myomonitor® |
|
The Myomonitor IV Analog Output System enables the integration of 16 full-bandwidth data channels from the Myomonitor IV Wireless EMG System into any analog data acquisition system. It uses a precision waveform generator with a low-noise filter to recreate the high-fidelity signals detected by Delsys Sensors.
|
 |
|
Specifically designed for:
- Gait Laboratories
- Biomechanics Laboratories
- Exercise Physiology Research
- Ergonomic Studies
- Robotics & HCID Interfacing
| Features:
- 16 channels, 4 kHz sampling
- BNC & Mass terminations
- +/- 5 V Range
- Low noise, full signal conditioning
- 50 ms monitored latency
|
| |
 |
| |
 |
|
Typical Laboratory Setup |
 |
|
Analog Output System Components |
 |
|
Analog Filter Module Specifications*
| Filter Topology |
Butterworth, 8 pole |
| Bandwidth |
DC-500 Hz, +/- 10% |
| Voltage Range |
+/- 5 V |
| Setting Time |
<250 us |
| Baseline Noise |
0.250 mV (RMS,RTO) |
| Input: Output Error (SNR) |
<2% |
| Power Supply |
9 VDC |
| Dimensions |
205 mm x 108 mm x 57 mm |
| Mass |
300 g |
| Output Connections |
BNC, SCSI-68 (National Instruments) |
| Output Impedance |
<50mΩ ,buffered |
|
|
NI PCI-6723 Waveform Generator Specifications
| Output Range |
+/- 5 V |
| Resolution |
13 Bits |
| Number of Channels |
16 |
| Update Rate |
4000 samples/sec |
| PC Bus Requirement |
PCI |
| Dimensions |
174 mm x98 mm |
| Operating Temperature |
6-50 °C |
| Humidity |
5-90% RH, non-condensing |
| Output Connections |
VHDCI (National Instruments) |
|
| |
*All specifications are subject to change without notice. Specifications include the software configuration of the device. |