How is a ‘Cold Reset’ performed on the Myomonitor IV device?

March 4th, 2014|

Depress and Hold the ‘ON/OFF’ button on the Myomonitor. While holding this button in a depressed state, use the stylus to depress the ‘RESET’ button for at least 2 seconds. Release the ‘RESET’ button, and continue to hold the ‘ON/OFF’ [...]

The Myomonitor screen is on but the device never boots. What is the cause?

March 4th, 2014|

The most likely cause is a wrong power supply. It is important that you only use the power supply supplied with the Myomonitor, since any other power supply may damage the device. The Myomonitor power supply is manufactured by "Phihong" [...]

How can I integrate signals other than EMG into Myomonitor® devices?

March 4th, 2014|

There are three methods available for integrating additional external signals to Myomonitor recordings. 1) The preferred method is to use a Delsys Biosignal Sensor, than can connect directly to any available channel Myomonitor. Contact Delsys for custom modifications and/or custom [...]

I have a PC and my own A/D card. Can I use this with Delsys equipment?

March 4th, 2014|

All Delsys products that produce an analog signal output can be connected to any A/D card that can support +/-5V analog signals and is controlled by appropriate software. However, when using Delsys software (EMGworks®), it is only possible to use [...]

Can I use stimulation equipment while recording with Delsys EMG equipment?

March 4th, 2014|

Delsys devices are not designed to be used with stimulation devices. The fundamental performance issue is the amplitude of the detected stimulus as recorded by the DE-2.1 sensor. An amplitude at the recording site that exceeds +/-5mV will saturate the [...]

The Bagnoli™ Main Amplifier turns on but no signal is received. What are the possible causes?

March 4th, 2014|

The most likely cause is a problem with the Input Module cable. Check the cable carefully, making certain that it is not damaged or excessively twisted. Additionally check the cable connectors to ensure that the contacts and connector shell are [...]

Can I use Delsys systems on animals?

March 4th, 2014|

Yes, many studies and applications have successfully employed Delsys EMG sensors and systems on animals. Be sure correctly identify appropriate muscle sites and remove all fur/hair prior to applying the sensors. Wipe the site carefully with isopropyl alcohol to remove [...]

Can I use Delsys Equipment inside MRI machines?

March 4th, 2014|

Delsys EMG systems are not approved for use in MRI machines. MRI machines make use of very strong magnetic fields that may induce unwanted currents in metallic objects they traverse. The sensitive electronic components used in Delsys EMG systems will [...]