The Trigno™ Snap Lead Sensor can connect to industry-standard pinch electrodes, for a more “traditional” approach to sEMG detection. The sensor’s flexible clamp leads allow users to adjust their inter-electrode spacing as they deem suitable, and also enable data collection from difficult muscle sites.

Specifications
Quote Request
Trigno Snap Lead Sensor

Full EMG bandwidth

Built-in accelerometer

Robust wireless transmission

Compatible with standard snap electrodes

Trigno Snap Lead Sensor

Specifications

Trigno
User’s Guide

Typical Operating Range
40 m

RF Frequency Band
2400-2483 MHz (ISM band)

Power Consumption
<65 mW

Effective Radiated Power
9 mW

RF Protocol
Proprietary

EMG Channel Bandwidth
20 ± 5 Hz, >40 dB/dec
450 ± 50 Hz, >80 dB/dec

EMG Channel Sampling Rate
1925.93 samples/sec

EMG Channel Resolution Depth
20 ± 5 Hz > 40 dB/dec
450 ± 50 Hz > 80 dB/dec

EMG Channel Noise
<500 nV (rms), r.t.i.

ACC Channels Bandwidths
DC – 50 Hz ± 5 Hz, 20 dB/dec

Accelerometer Sampling Rate
148.1 samples/sec/axis

Accelerometer Resolution Depth
0.016 ± 0.001 g/bit (at ± 1.5 g)
0.063 ± 0.005 g/bit (at ± 6 g)

Accelerometer Noise
0.004g (rms, ± 1.5g)